![Hach DR 6000 User Manual Download Page 72](http://html1.mh-extra.com/html/hach/dr-6000/dr-6000_user-manual_576875072.webp)
72
Standard programs
Table 5 Barcode program options
Options
Description
More
For further Options
Save icon
Depending on setup under
Instrument Setup > Data Log Setup > Auto Store: On
or
Off
the
stored data is called up with the
Save
symbol or a manual save is performed.
Abs % Trans
Switches from % transmittance to absorbance in reading mode.
Send Data icon
To send data to a printer, computer or USB memory stick (USB A) or network.
Timer icon
This functions as a stopwatch. It helps to make sure that the steps of an analysis are correctly
timed (e. g. reaction times, wait times, etc., can be exactly specified). When the specified time
has elapsed, a sound is emitted. The use of the timer has no influence on the reading program.
Dilution Factor Off/On
A corrective dilution factor can be entered in order to take account of certain properties.
The number entered at the dilution factor prompt will be multiplied by the result to compensate
for the adjustment. For example, if the sample has been diluted by a factor of 2, enter 2. The
default setting of the dilution factor is turned off.
Note: During use of a dilution factor, the dilution symbol is shown on the display.
Note: If undiluted samples are used, set the dilution factor off.
Chemical Form
Some of the stored tests/methods allow to select the chemical form and the measuring range.
Reagent Blank: Off/On
The function "Reagent Blank" is available for several barcode tests. This enables the reagent
blank value to be added to or subtracted from, the subsequent readings. The reagent blank
value shifts the calibration curve along the y-axis, without changing the shape or gradient of the
curve. The effect corresponds to a y-axis intercept of the calibration straight line. This is clarified
by the following formula: concentration =[conc. factor) * Abs] - (reagent blank)
Edit
To modify an existing program
Lot information
All relevant manufacturing data is stored here.
Save as User Program
To store the selected parameters as a User Program, refer to
.
Multiple determination
This option allows multiple measurements of the same sample.
Reading Mode
Barcode test function not available.
Recall measurement
data
Recall saved measurement data, wave-length scans or time courses, see
.
Instrument Setup mode
For basic instrument data, refer to
Summary of Contents for DR 6000
Page 2: ......
Page 16: ...16 General information ...
Page 132: ...132 Expanded programs ...
Page 142: ...142 Troubleshooting ...
Page 145: ...145 Section 10 Contact ...
Page 146: ...146 Contact ...
Page 152: ...152 Help Guide ...
Page 156: ...156 Index ...