Expanded programs
Process Timer key
1.
Tap on the
Process Timer
key to enter wait-, reaction- or
handling times.
2.
Tap on
OK
and confirm the entry.
Note: This time is integrated into the measurement process.
Note: The entered reading sequence is displayed.
[R] key / Reading...
1.
Press
[R]
to program a reading of the substance that is to
be analyzed.
2.
Press
Reading...
and enter the wavelength to be used for
the measurement.
3.
If readings are to be completed at a number of wavelengths,
repeat the steps starting with 1 for each wavelength.
Note: The entered reading sequence is displayed.
Key { }
Elements of the reading sequence that are to be repeated are
placed in brackets.
The left bracket "{"marks the start of the sequence that is to be
repeated and the right bracket "}" marks the end.
Note: The button with the right parenthesis remains inactive until a left
bracket is entered.
1.
To start the repetition sequence tap on
{
.
2.
Tap on the key that defines the sequence to be repeated:
[Z]
or
[R]
.
3.
Tap on the key
Zeroing...
or
Reading...
and enter the
wavelength to be used for the measurement. Define
additional steps that are to be included in the repetition
sequence.
4.
Press
}
.
5.
Once the measurement process has been defined, click
Next
.
Alternate entry of measurement process via function list
A pre-programmed selection can be accessed and edited via
List
.