21
Installation
Figure 3
Cell compartments
3.6.2
How to position the universal cell adapter
1.
Open the cell compartment.
2.
Raise the universal cell adapter by approx. 1 cm.
3.
Rotate the universal cell adapter such that the guide of the
required cell profile faces left toward the cell compartment
(1).
4.
Press the universal cell adapter down until it snaps into
place.
Figure 4
How to position the universal cell adapter
1
Cell compartment for round cells (1)
2
Cell compartment with universal cell adapter (2)