66
Standard programs
Maximum Method Volume/Sample volume
1.
Analyze a sample without added standard solution in
accordance with the instructions in the Procedures Manual.
When the reading is complete, leave the sample cuvette/cell
in the cuvette/sample cell compartment.
2.
Press
Options > More > Standard Addition
.
An overview of the data of the standard addition procedure is
displayed.
3.
Press
OK
to accept the standard values for standard
concentration, sample volume (total) and standard addition
volume.
4.
Press
Edit
, to change these values.
5.
Press the key of the value to change it. Use the
alphanumeric keypad to change the value.
6.
Press
Next
to confirm.
7.
Press the keys to change the standard addition volumes.
Use the alphanumeric keypad to enter the new data.
8.
Press
OK
to confirm.