![Hach DR 6000 User Manual Download Page 104](http://html1.mh-extra.com/html/hach/dr-6000/dr-6000_user-manual_576875104.webp)
104
Expanded programs
The following formulas are available:
K
1
A
1
+ K
2
A
2
K
1
A
1
+ K
2
A
2
+ K
3
A
3
K
1
A
1
+ K
2
A
2
+ K
3
A
3
+ K
4
A
4
K
1
A
1
/ K
2
A
2
(K
1
A
1
+ K
2
A
2
) / K
3
A
3
(K
1
A
1
+ K
2
A
2
) / (K
3
A
3
+ K
4
A
4
)
A
1
refers to the absorbance at wavelength 1
A
2
refers to the absorbance at wavelength 2, etc.
K
1
refers to the coefficient at wavelength 1
K
2
refers to the coefficient at wavelength 2, etc.
If a subtraction has to be completed, the factors can be entered
with a minus sign.
6.
To change the wavelength, press the
x
key.
7.
Enter the required wavelength factor using the alphanumeric
keypad.
8.
Press
OK
to confirm the selection.
9.
To change a coefficient, press one of the "
K
X
" keys.
10.
Enter the required coefficient using the alphanumeric
keypad.
11.
Press
OK
to confirm the selection.
Note: Up to 5-digit numeric values can be entered. A maximum of
four decimals places are possible.
Concentration Factor:
1.
Press
Concentration Factor: Off
in the "Options" menu.
2.
Press
On
to highlight this feature.
3.
Press
Factor
.
4.
Enter the factor by which the absorbance values are to be
multiplied.
5.
Press
Unit
.
6.
Select the respective concentration unit from the displayed
list.
Summary of Contents for DR 6000
Page 2: ......
Page 16: ...16 General information ...
Page 132: ...132 Expanded programs ...
Page 142: ...142 Troubleshooting ...
Page 145: ...145 Section 10 Contact ...
Page 146: ...146 Contact ...
Page 152: ...152 Help Guide ...
Page 156: ...156 Index ...