![Hach DR 6000 User Manual Download Page 74](http://html1.mh-extra.com/html/hach/dr-6000/dr-6000_user-manual_576875074.webp)
74
Standard programs
5.5.5
Update/edit barcode tests
5.5.5.1
Update barcode tests with test packaging (not available on all models)
Using the data provided in the barcode, the instrument
automatically sets the reading wavelength and factors. If a
deviation in barcode data from the stored data is detected, the
instrument will require a data update and display the error
message "Barcode control number? Update program data!"
1.
Hold the packaging of the barcode test with the RFID symbol
in front of the RFID module of the instrument.
A sound indicates the successful transfer of data; the test is
automatically assessed.
If the packaging is held in front of the RFID module a second
time, the current lot information of the test is shown in the
display. The test can be performed next according to the
process specification.
5.5.5.2
Manual update of a barcode test
Note: Only in very few cases does the revision of a test require all test
specifications to be updated.
The various data types that determine a barcode test are
described in
Table 6
.
Table 6 Program update options
Program point
Description
Name
Name of the analysis parameter
Version
An abbreviation assigned by the user or the version number, etc. is entered here
Measurement process
Exact definition of the test: the number of wavelengths at which readings are carried out, the
number of absorbance readings needed, the keys to be used, any waiting periods between
readings, etc.