82
Expanded programs
Chemical form:
1.
Enter the chemical formula used in the display to represent
the analysis parameter.
2.
Press
Continue
.
Note: For subsequent programming steps, refer to
onwards.
6.1.1.2
Multi wavelength settings
When selecting the Multi-Wavelength mode, the following
parameters can be defined:
Units:
1.
Select the required unit from the list
2.
Press
Continue
.
Note: User-specific units not included in this list can be added in the edit
program under Program Options, Edit. Select Units, Edit and then
New.
Absorbance formula (Multi Wavelength program type):
Define the wavelengths and the coefficients used in the formula.
The absorbance formula defines the calculation for the multi
wavelength reading.
1.
Press
Formula
.
2.
Select the formula relevant to the programming from the
displayed list.
3.
Press
OK
to confirm the selection.
List of the available absorbance formulae
A
1
refers to the absorbance at wavelength 1
A
2
refers to the absorbance at wavelength 2, etc.
K
1
refers to the coefficient at wavelength 1
K
2
refers to the coefficient at wavelength 2, etc.
If a subtraction has to be performed, the factors can be entered
with a minus sign.