Expanded programs
When results exceed allowable tolerances (given in the quality
control certificate to the test record), contact the manufacturer.
1.
Press
Verification Kit
in the "Optical Checks" menu.
2.
Press
Nominal Values
.
3.
Press
Edit
.
An automatic menu guidance queries values (filters,
wavelength, nominal values and tolerances) given in the
quality control certificate, to the following specifications:
•
Stray light
•
Photometrical accuracy
•
Wavelength accuracy
4.
Press
OK
when all values are entered and the overview is
displayed.
5.
Press
Verification
.
6.
Insert the universal cell adapter(
) into the
cell compartment (2).
7.
Remove any cells from the cell compartment.
8.
Close the cell compartment cover.
9.
Press
Start
.