![Hach DR 6000 User Manual Download Page 127](http://html1.mh-extra.com/html/hach/dr-6000/dr-6000_user-manual_576875127.webp)
127
Expanded programs
11.
Press
Options
to set further functions:
New
: Defines an additional pipette.
Delete
: Deletes a defined pipette.
Edit
: Modifies settings for a defined pipette.
Reminder
: Defines a reminder interval to determine when
the user is prompted to return to this AQA measurement.
History
: List of all previously performed pipette checks.
12.
Press
Close
to exit the options menu.
6.7.5
Instrument backup
Prior to the next service date, the "
Instrument Backup
" offers
the opportunity to store all programs, measurement data, user
IDs, passwords, as well as all adjustable data on a USB stick.
1.
Press
Instrument Backup
in the "System Check" menu.
2.
Insert the USB stick (refer to
).
3.
Press
Store
to start a Backup.
If the file was stored, the message "Instrument Backup is stored"
will be displayed.
4.
Press
OK
to return to the "System Check" menu.
Restore backup data:
Note: All current data will be overwritten when restoring the Backup file!
1.
Press
Instrument Backup
in the "System Check" menu.
2.
Insert the USB stick with the backup (refer to
3.
Press
Restore
to pass back the data.