![Hach DR 6000 User Manual Download Page 79](http://html1.mh-extra.com/html/hach/dr-6000/dr-6000_user-manual_576875079.webp)
79
Section 6
Expanded programs
6.1
User programs
User programs provide the opportunity to complete "made to
measure" analysis.
The "User Programs" database is empty when the instrument
leaves the factory and is used to accommodate programs created
by users for specific needs. User programs with 13 mm round
cells can only be performed with the empty cells LCW906. Here
are a few examples of entries:
•
Programming of user-created procedures.
The analysis procedure must be developed first, before it can
be programmed. The user must define or determine the
program sequences, absorbance formulas, reading
wavelengths, factors, measuring range limits, etc.
•
Modified tests
•
Assignment of user programs to the "Favorites" menu for
frequently used tests.
•
Creation of a specific selection of methods and tests.
1.
Select
User Programs
in the "Main Menu"
2.
Select
Program Options
.
The
Program Options
menu contains several input and editing
options (
Table 7
):
Table 7 Program options for user programs
Options
Description
New
Select
New
to program a new user program.
Note: The first time Program Options is selected, only the New option is available. The other
options remain inactive (grey) until the first program has been created.
Add to favorites
Select
Add to favorites
to add an existing user program to the Favorites list.
Edit
Select
Edit
to modify an existing program
Delete
Select
Delete
to remove a program from the list of user programs. The program will be
simultaneously deleted from the Favorites list.
Summary of Contents for DR 6000
Page 2: ......
Page 16: ...16 General information ...
Page 132: ...132 Expanded programs ...
Page 142: ...142 Troubleshooting ...
Page 145: ...145 Section 10 Contact ...
Page 146: ...146 Contact ...
Page 152: ...152 Help Guide ...
Page 156: ...156 Index ...