60
Standard programs
1.
Press the
Timer
icon.
2.
Press
OK
to start the first timer.
The timer will count down on the screen.
3.
Press the
Timer
symbol and confirm with
OK
to start the
next activity with predetermined time for the stored program.
Note: Press Close to view the reading screen while the timer is
running. The time will be shown on the lower left side instead of the
date.
Note: Press Cancel to stop the timer midway through the
countdown.
Note: The timer will beep when the timed interval ends.
A general purpose timer is also available in many programs. Tap
on the symbol for the
timer
and select
General Timer
. Enter the
length of the timed interval and press
OK
to start the timer. The
timer will beep when the timed interval ends.
5.4.4
Set the dilution factor
The Dilution Factor function is used to calculate the original
concentration of a sample that has been diluted by a known ratio.
1.
Press
Options
>
More… > Dilution Factor
.
The number entered at the dilution factor prompt will be
multiplied by the result to compensate for the adjustment.
For example , if the sample has been diluted by a factor of 2,
enter 2. The default setting of the dilution factor is turned off.
2.
Press
OK
to confirm. Press again
OK
.
Note: During use of a dilution factor, the dilution symbol is shown on the
display.
Note: If using undiluted samples, set the dilution factor off again.