Expanded programs
6.7.3.3
Stray Light Check
With the stray light check, the stray light in the device is
measured at 340 nm.
1.
Press
Stray Light Check
in the "Optical Checks" menu.
2.
Remove any present cells from the cell compartment.
3.
Insert the universal cell adapter for 10-mm rectangular cells
into the cell compartment (2).
4.
Close the cell compartment cover.
5.
Press
Zero
.
6.
Insert the analysis cell into the cell compartment (2).
7.
Close the cell compartment cover.
8.
Press
Start
.
The mean value is calculated from three consecutive
absorbance measurements.
The result will be displayed.
9.
Compare the result with the required value from the analysis
cell.
10.
Press
Cancel
to return to Optical Checks.
6.7.3.4
Absorbance Check
With the absorbance check the photometric accuracy and
repeatability of the device is checked.
1.
Press
Absorbance Check
in the "Optical Checks" menu.
2.
Press
to enter the wavelength.
3.
Press
OK
to confirm.
4.
Remove any cells from the cell compartment.
5.
Close the cell compartment cover.
6.
Press
Zero
.