Section 8-20 - Replacement of the Voluson Station Components
8-20-21
Replacement of the Back Handle complete
8-20-21-1
Manpower
One person, 10 minutes
8-20-21-2
Tools
Phillips screwdriver 1
8-20-21-3
Preparations
1.) Power Off/Shutdown the system as described in
Section 3-6-3 on page 3-72
.
2.) Remove the Patient Monitor as described in
Section 8-20-10 on page 8-70
.
8-20-21-4
Back Handle - Removal Procedure
1.) Screw off the 5 screws and carefully remove the Top and Bottom Back Handle.
NOTE:
Do not lose the Patient Monitor Adapter !
8-20-21-5
Back Handle - Installation Procedure
1.) Place the new Back Handle on its original position and also the Patient Monitor Adapter.
2.) Fasten it with the 5 screws.
3.) Mount the Patient Monitor as described in
Section 8-20-10-5 on page 8-70
.
Figure 8-103 screw off the 5 screws
Patient
Monitor
Adapter
Patient
Monitor
Adapter
short screw