Section 3-6 - Completing the Set Up
3-6-3
Power Off/ Shutdown
•
Standby Mode
NOTE:
The Standby Mode on the Voluson i / Voluson e is available with Software 7.x.x. and higher.
1.) If not already in read mode, freeze the image.
2.) Press the
On/Off
button (see:
Figure 3-87
) on the control panel. Following dialog appears.
3.) Select the STANDBY button.
The system performs an automatic shutdown sequence and will remain in
Standby
mode.
4.) Leave the main power cable connected to keep the standby battery fully charged.
NOTE:
Once the lid of the Voluson i / Voluson e system is closed completely, Standby mode is activated.
NOTE:
It is recommended to perform a complete system power down (see:
• Full Shutdown
) at least once a
week.
•
Full Shutdown
1.) If not already in read mode, freeze the image.
2.) Press the
On/Off
button (refer to:
Figure 3-87
). A dialog window (see:
Figure 3-90
, above) appears.
3.) Select the SHUTDOWN button. The system performs an automatic full shutdown sequence.
4.) After complete power down, unplug the main power cable from the AC wall outlet socket.
5.) Unplug the main power cable connector from the AC input socket on the Voluson Station / Cart or the
AC-DC converter of the system.
NOTE:
Connect the power cable to the main power outlet again as soon as possible and at least before you
switch on the main power and press the On/Off button.
Figure 3-90 Shutdown, Restart or Standby
NOTICE
!! NOTICE:
The battery is automatically charged when the system is plugged into an active wall outlet.
It takes approximately 10 hours to charge a flat battery.
If the system is left in the OFF condition (system unplugged from the wall outlet) for an extended period
of time (3 to 5 days or more), the system may not boot up or may beep when turned ON. If this occurs,
the system needs to recharge the batteries. This may take 15 minutes to as long as 10 hours,
depending on battery age, system input voltage and system temperature. To avoid draining the battery,
the system should remain plugged into the wall outlet with the wall outlet power switch in ON position.
NOTICE
!! NOTICE:
After turning off a system, wait at least ten seconds before turning it on again.
The system may not be able to boot if power is recycled too quickly.
NOTICE
!! NOTICE:
If the system is still running when disconnecting the main power, the internal power supply will switch
over to the battery power. The internal battery supplies the system with power until it is drained to 30%
capacity (invoking shutdown), or shutdown has been invoked by the user.
NOTICE
!! NOTICE:
After turning off a system, wait at least ten seconds before turning it on again.
The system may not be able to boot if power is recycled too quickly.
Summary of Contents for Voluson i BT06
Page 2: ......
Page 11: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL ix ZH CN KO...
Page 44: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL xlii Table of Contents...
Page 514: ...GE VOLUSON i VOLUSON e DIRECTION KTI106052 REVISION 10 SERVICE MANUAL IV Index...
Page 515: ......