Chapter 7 - Diagnostics/Troubleshooting
7-15
Section 7-6 How to use the Auto Tester program
(cont’d)
7.) Activate the “Auto Tester” program by clicking the “Record” icon on the displayed screen.
8.) Start scanning.
You can scan normally and everything will be recorded to the program (up to several hours).
NOTE:
It is important that you are recording the processes where the errors normally occur.
The following screen will appear.
Figure 7-15 Start Auto Tester
Figure 7-16 Stop Autotester
Figure 7-17 Auto Tester Finished
“Record”
icon
Stop the program by clicking on [
Stop
] shown on the screen,
or by pressing the [
ALT GR
] key on the alphanumeric keyboard.
“
Disc
” icon