GE
V
OLUSON
i / V
OLUSON
e
D
IRECTION
KTI106052, R
EVISION
10
S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-53
8-19-9
Replacement of the Mounting Platform
8-19-9-1
Manpower
One person, 20 minutes
8-19-9-2
Tools
Phillips screwdriver 1
Torx screwdriver TX8
8-19-9-3
Preparation
1.) Power OFF/Shutdown the system as described in
Section 3-6-3 on page 3-72
.
2.) Disconnect and remove the Voluson i / Voluson e.
3.) Remove the Cover Bottom as described in
Section 8-19-4 on page 8-48
.
4.) Remove the Back Column Cover as described in
Section 8-19-5 on page 8-49
.
8-19-9-4
Mounting Platform - Removal Procedure
1.) Screw off the screw.
2.) Screw off the 4 Torx screws.
8-19-9-5
Mounting Platform - Installation Procedure
1.) Place the new Mounting Platform and fasten it with the screws.
2.) Mount the Back Column Cover as described in
Section 8-19-5-5 on page 8-49
.
3.) Mount the Cover Bottom as described in
Section 8-19-4-5 on page 8-48
.
Figure 8-72 unscrew the screw
Figure 8-73 unscrew the Torx screws
Summary of Contents for Voluson i BT06