Chapter 3. Making Measurements
74
USM Go Operator’s Manual
3.2.1
Setting the Angle Beam Probe Parameters (cont.)
3.
Use the joystick (
) to select the
THICKNESS
function, and input the thickness of the test piece. The thickness of a solid rod
should be input as 1/2 of its diameter.
4.
Use the joystick (
) to select the
X VALUE
function, and input the user-determined X value for the probe. This compensates for
the distance from the BIP to the front of the probe wedge.
5.
Use the joystick (
) to select the
O DIAMETER
function, and input the outside diameter (from 50 to 2000 mm) of a curved test
piece. When set to
FLAT
, curvature correction is not applied.
Note:
Based on the values input for the
PROBE ANGLE
,
THICKNESS
,
X VALUE
, and
O DIAMETER
parameters, the instrument
operates in one of five modes. Each of these modes is represented by an icon in the
icon area
at the upper left corner of the screen.
See
Figure 6 on page 17
for an illustration of these icons.