A.8 Protection
Damp Heat & Humidity (Storage):
10 Cycles:
10 hr at +60°C down to +30°C, 10 hr at +30°C up to +60°C,
Transition within 2 hr (507.4)
Temperature Shock (Storage):
3 Cycles:
4 hr at -20°C up to +60°C, 4 hr at +60°C,
Transition within 5 minutes (503.4 Procedure II)
Vibration:
General Exposure:
1 hr each axis, 514.5-5 Procedure I, Annex C, Figure 6
Shock:
6 Cycles each axis
, 15 g, 11 ms, half-sine (516.5 Procedure I)
Loose Cargo (in Shipping Container):
514.5 Procedure II
Transit Drop (Packaged for Shipment): 26 Drops
, 516.5 Procedure IV
Dust Proof/Dripping Water Proof:
As per the IEC 529 specification for IP67 classification
Operating Temperature Range:
0 to 55°C
Storage Temperature Range:
-20 to +60°C, 24 hr with battery