![GE Ultrasonic USM Go Operator'S Manual Download Page 112](http://html.mh-extra.com/html/ge/ultrasonic-usm-go/ultrasonic-usm-go_operators-manual_128490112.webp)
Chapter 3. Making Measurements
94
USM Go Operator’s Manual
3.13 DGS Evaluation Mode (cont.)
In addition to the influencing variables already mentioned, these other factors affect the curve shape:
•
sound attenuation
•
transfer losses
•
amplitude correction value
•
probe
Also, the following probe parameters affect the curve shape:
•
element or crystal diameter
•
frequency
•
delay length
•
delay velocity
You can adjust these parameters in such a way that you can use the
DGS
method with many different probes and on various materials.