Chapter 2. Instrument Setup
USM Go Operator’s Manual
59
2.7 Using the Calibration Reminder Alarm
The USM Go incorporates a timed alarm feature that causes an icon to appear on user-defined input intervals between 0.5 and 4.0 hours. To
use this alarm:
1.
In the
SETUP
menu, activate the
CONFIG2
submenu using the joystick (
).
2.
Use the joystick (
) to select the function titled
CAL REMINDER
, then press the center of the joystick (
) to activate the
function. Input the intervals at which the reminder alarm will trigger (between 0.5 and 4.0 hours). Setting this value automatically resets
the alarm to trigger on the specified intervals. This function also allows the alarm to be disabled.
3.
Press the function toggle or move the joystick (
) to change the alarm interval.
4.
Select and activate
CAL RESET
to acknowledge the triggered alarm and reset it to resume normal operation.
Note:
The selection of the
PROBE DELAY
or
VELOCITY
functions also resets the alarm.