Chapter 2. Instrument Setup
USM Go Operator’s Manual
55
2.6.2 Using AUTOCAL
Note:
While following the instructions in this section, refer to
Figure 14 on page 56
.
1.
In the
ACQUIRE
menu, activate the
AUTOCAL
submenu using the joystick (
).
2.
Use the joystick (
) to select the function titled
S-REF1
. Then, press the function toggle or move the joystick (
) to change
the value to match the thickness of the
thinner
calibration standard. The S-REF1 function has both coarse and fine adjustment modes.
Coarse adjustments are made with the function toggle, while fine adjustments are made with the joystick.
3.
Use the joystick (
) to select the function titled
S-REF2
. Then, press the function toggle or move the joystick (
) to change
the value to match the thickness of the
thicker
calibration standard. The S-REF2 function has both coarse and fine adjustment modes.
Coarse adjustments are made with the function toggle, while fine adjustments are made with the joystick.
4.
Apply couplant and couple the probe to the
thinner
calibration standard. With the
A START
function highlighted, press the function
toggle or move the joystick (
) to shift the starting point of Gate A until Gate A lies over the echo corresponding to the thickness of
the thinner standard (see
Figure 14 on page 56
). Coarse adjustments are made with the function toggle, while fine adjustments are made
with the joystick.
Summary of Contents for Ultrasonic USM Go
Page 2: ......
Page 4: ... no content intended for this page proceed to next page ...
Page 18: ...Preface xvi USM Go Operator s Manual no content intended for this page proceed to next page ...
Page 162: ...144 USM Go Operator s Manual no content intended for this page proceed to next page ...
Page 178: ...Appendix C EN 12668 Specifications 160 USM Go Operator s Manual ...
Page 189: ......