Chapter 3. Making Measurements
110
USM Go Operator’s Manual
3.16.2b Defining a BOLD LINE Measurement Reference
(JISDAC-SETUP-BOLD LINE)
The user-defined measurement reference is selected as follows:
1.
Access the
SETUP
submenu in the
JISDAC
menu.
2.
Use the
BOLD LINE
function to select the H, M, or L line.
3.
Note that the selected line is now shown in bold on the A-Scan display. This line now serves as the reference level from which all
reference measurements (such as A%rA or dBrB) are made.
3.16.2c Setting TRANSFER CORRection
(JISDAC-MAT ATTN-TRANSFER CORR)
The
TRANSFER CORR
function applies dB compensation for the difference in coupling conditions between the known standard and the
test piece.
1.
Access the
MAT ATTN
submenu in the
JISDAC
menu.
2.
Adjust the
TRANSFER CORR
function as required to compensate for coupling differences.
Note:
When
TRANSFER CORR
is set to any value other than 0, the displayed gain value appears in a contrasting color, indicating that
the amount of applied gain differs from the listed
“instrument”
gain.
Summary of Contents for Ultrasonic USM Go
Page 2: ......
Page 4: ... no content intended for this page proceed to next page ...
Page 18: ...Preface xvi USM Go Operator s Manual no content intended for this page proceed to next page ...
Page 162: ...144 USM Go Operator s Manual no content intended for this page proceed to next page ...
Page 178: ...Appendix C EN 12668 Specifications 160 USM Go Operator s Manual ...
Page 189: ......