Chapter 3. Making Measurements
98
USM Go Operator’s Manual
3.13.2 Record the Reference Echo that Defines the DGS Curve (cont.)
Note:
The
AUTO 80
function automatically sets the Gate A triggering echo to 80% of full screen height.
Note:
Only one
DGS
reference echo can be stored at a time. To delete the currently stored reference, access the
REF CORR
submenu,
select
DELETE REF
, and follow the on-screen prompts.
Note:
When a
DGS
reference echo is stored, two measurement result boxes are automatically set, if not already configured, to display
SA
and
A%A
values.
IMPORTANT:
The following two adjustments should be made prior to recording a reference echo. Changing these values after the
DGS
curve is generated will cause the curve to be deleted.
•
REF ATTEN
(found in the
REF CORR
submenu) - Specify a sound attenuation value, in dB per inch or mm of material thickness, for the
material from which the known standard is made.
•
AMPL CORRECT
(found in the
REF CORR
submenu) - Correction required when using an angle-beam type probe. This value is
specified on the probe data sheet.
Summary of Contents for Ultrasonic USM Go
Page 2: ......
Page 4: ... no content intended for this page proceed to next page ...
Page 18: ...Preface xvi USM Go Operator s Manual no content intended for this page proceed to next page ...
Page 162: ...144 USM Go Operator s Manual no content intended for this page proceed to next page ...
Page 178: ...Appendix C EN 12668 Specifications 160 USM Go Operator s Manual ...
Page 189: ......