2.4.7 Using the Phantom PRF Feature
When activated, this diagnostic feature varies the PRF to identify any wrap-around signals, which are phantom echoes caused by a PRF
setting that is too high. When this feature is activated, the time-based position of the phantom echoes varies while the true echoes remain
stationary on the display screen.
1.
In the
SETUP
menu, activate the
CONFIG2
submenu using the joystick (
).
2.
Use the joystick (
) to select the function titled
PHANTOM PRF
. Press the joystick (
) to activate this function.
3.
Press the function toggle or move the joystick (
) to choose either
ON
or
OFF
.
4.
After making your choice, press the center of the joystick (
) to deactivate the function.