Chapter 3. Making Measurements
102
USM Go Operator’s Manual
3.14 dB REF Evaluation Mode
Note:
Evaluation modes, such as
dB REF
, are selected via the
EVAL MODE
function in the
EVAL
menu. The selected evaluation mode
menu then appears in the
ACQUIRE
menu bar.
When
dB REF
is activated, the amplitude of the highest echo in A-Gate becomes the reference echo against which subsequent echo
amplitudes are evaluated. At the time of the
dB REF
activation, the gain setting also becomes a reference against which subsequent gain
values are compared.
1.
Access the
SETUP
submenu in the
dB REF
menu.
2.
Locate the reference echo in Gate A and adjust the gain as required. Then, use the
RECORD
function to record the reference echo. The
function box now indicates that a reference has been stored. Also note the following:
•
The instrument gain and the reference echo amplitude are now shown in the function box titled
REFERENCE
.
•
The instrument automatically switches the
dB REF MODE
function to
ON
, as indicated on the display.
•
Setting
MODE
to
OFF
allows you to disable the
dB REF
mode without deleting the reference echo.
•
Use
DELETE REF
function to delete the current reference.
Summary of Contents for Ultrasonic USM Go
Page 2: ......
Page 4: ... no content intended for this page proceed to next page ...
Page 18: ...Preface xvi USM Go Operator s Manual no content intended for this page proceed to next page ...
Page 162: ...144 USM Go Operator s Manual no content intended for this page proceed to next page ...
Page 178: ...Appendix C EN 12668 Specifications 160 USM Go Operator s Manual ...
Page 189: ......