![GE Ultrasonic USM Go Operator'S Manual Download Page 88](http://html.mh-extra.com/html/ge/ultrasonic-usm-go/ultrasonic-usm-go_operators-manual_128490088.webp)
Chapter 3. Making Measurements
70
USM Go Operator’s Manual
3.1.3a Defining Gate-Alarm Logic (SETUP-CONFIG2-GATE A or B LOGIC)
Each gate alarm can be triggered by one of two circumstances: either when an A-Scan echo crosses the gate or when no A-Scan echo crosses
the gate. Use the following procedure to specify the gate
LOGIC
settings:
1.
In the
SETUP
menu, activate the
CONFIG2
submenu using the joystick (
).
2.
Use the joystick (
) to select the function titled
GATE A
(or
GATE B
)
LOGIC
. Press the center of the joystick (
) to activate
the function.
3.
To choose the gate-alarm triggering logic, move the joystick up or down (
). The following options are available:
•
POSITIVE
- An A-Scan signal crosses the gate
•
NEGATIVE
- No A-Scan signal crosses the gate
•
OFF
- No alarm is connected to the selected gate
Note:
Gate A and Gate B can be configured so that either gate triggers the alarm.
4.
After making your choice, press the center of the joystick (
) to deactivate the function.