Chapter 2. Instrument Setup
52
USM Go Operator’s Manual
2.5.1 Setting the A-Scan Range (cont.)
To set the A-Scan range:
1.
In the
ACQUIRE
menu, activate the
RANGE
submenu using the joystick (
).
2.
Use the joystick (
) to select the function titled
RANGE
, which has both coarse and fine adjustment modes. Coarse adjustments
are made with the
function toggle
, while fine adjustments are made with the
joystick
. When
“RANGE”
appears in all capital letters,
coarse adjustments are being made, while fine adjustments occur when
“range”
appears in all lower-case letters.
3.
Press the function toggle or move the joystick (
) to change the range setting. Values from 0.040 to 1100 in. are allowed.
Note:
Pressing both buttons on the
Function Toggle
simultaneously resets the
RANGE
to the default value of
10.000 in.
4.
After completing your selection, move the joystick up or down (
) to navigate away from this function.