Chapter 2. Instrument Setup
USM Go Operator’s Manual
49
2.4.8 Selecting a Rectification Mode (cont.)
1.
In the
ACQUIRE
menu, activate the
RECEIVER
submenu using the joystick (
).
2.
Use the joystick (
) to select the function titled
RECTIFY
. The following options are available:
•
NEG HALFWAVE
-
Shows the negative component of the RF signal but displays it in a positive orientation.
•
POS HALFWAVE
-
Shows the positive component of the RF signal.
•
FULLWAVE
-
Shows the positive and negative halves of the RF wave, but both are oriented in the positive direction.
•
RF
-
Shows the echo with no rectification.
3.
Press the function toggle or move the joystick (
) to select the desired rectification method.
4.
After completing your selection, move the joystick up or down (
) to navigate away from this function.