![GE Ultrasonic USM Go Operator'S Manual Download Page 72](http://html.mh-extra.com/html/ge/ultrasonic-usm-go/ultrasonic-usm-go_operators-manual_128490072.webp)
Chapter 2. Instrument Setup
54
USM Go Operator’s Manual
2.6 Calibrating the Instrument
To calibrate the USM Go, follow the instructions in this section.
2.6.1 Pre-Calibration Check List
To improve the accuracy and quality of your calibration, be sure that the following steps have been taken before beginning the calibration:
•
Install the probe
•
Adjust the
DUAL (RECEIVER)
setting to match probe
•
Set the material type
•
Set the
DISPLAY DELAY
to 0 (recommended)
•
Set the
PRF
to
AUTO LOW
•
Set
TCG
to
OFF
•
Set
REJECT
to 0 (recommended).
Summary of Contents for Ultrasonic USM Go
Page 2: ......
Page 4: ... no content intended for this page proceed to next page ...
Page 18: ...Preface xvi USM Go Operator s Manual no content intended for this page proceed to next page ...
Page 162: ...144 USM Go Operator s Manual no content intended for this page proceed to next page ...
Page 178: ...Appendix C EN 12668 Specifications 160 USM Go Operator s Manual ...
Page 189: ......