Chapter 3. Making Measurements
USM Go Operator’s Manual
71
3.1.3b Assigning Alarm Output Indicator Lights
(SETUP-CONFIG2-OUTPUT SELECT)
As an option, one reading box may be configured as a virtual LED (see
“Setting Gate Alarms and Outputs” on page 69
), which is green
when there is no fault condition and red when the alarm has been triggered. This virtual LED corresponds to an
OUTPUT
, which is in-turn
assigned to a gate alarm. When an alarm is triggered, the virtual LED is illuminated (except when the
GATE LOGIC
is set to
OFF
). Use the
following procedure to indicate which gate activates the virtual LED:
1.
In the
SETUP
menu, activate the
CONFIG2
submenu using the joystick (
).
2.
Use the joystick (
) to select the function titled
OUTPUT SELECT
. Press the center of the joystick (
) to activate the
function.
3.
To configure the gate-alarm virtual LED operation, move the joystick up or down (
). The following options are available:
•
A(+), A(-)
- the virtual LED indicates the Gate A alarm is triggered (see previous page for description of
+
and
-
alarms)
•
B(+), B(-)
- the virtual LED indicates the B Gate alarm is triggered (see previous page for description of
+
and
-
alarms)
•
A or B(+), A or B (-)
- the virtual LED indicates either Gate alarm is triggered (see previous page for description of
+
and
-
alarms)
4.
After making your choice, press the center of the joystick (
) to deactivate the function.
Summary of Contents for Ultrasonic USM Go
Page 2: ......
Page 4: ... no content intended for this page proceed to next page ...
Page 18: ...Preface xvi USM Go Operator s Manual no content intended for this page proceed to next page ...
Page 162: ...144 USM Go Operator s Manual no content intended for this page proceed to next page ...
Page 178: ...Appendix C EN 12668 Specifications 160 USM Go Operator s Manual ...
Page 189: ......