Chapter 3. Making Measurements
90
USM Go Operator’s Manual
3.10.2 Setting Transfer Correction (DAC/TCG-MAT ATTN-TRANSFER CORR)
To adjust the dB compensation for differences in ultrasonic coupling between the standard and the test piece:
1.
Access the
MAT ATTN
submenu in the
DAC/TCG
menu.
2.
Adjust the
TRANSFER CORR
function as required to compensate for coupling differences.
Note:
When
TRANSFER CORR
is set to any value other than 0, the displayed gain value appears in a contrasting color, indicating that
the amount of applied gain differs from the listed
“instrument”
gain.