Chapter 3. Making Measurements
96
USM Go Operator’s Manual
3.13.1 Specifying a Probe and Preparing to Record the Reference Echo (cont.)
3.
If the user-defined probe type (
PROBE #0
) is selected, you can then select the
PROBE NAME
function. Then use the
Gain Toggle
and the
Function Toggle
to input a new name. Note that selecting any
PROBE #
value other than 0 will prevent you from modifying
the
PROBE NAME
or any of the other settings described in this section.
4.
If the user-defined probe type (
PROBE #0
) is selected, you must then select the
DGS PROBE
submenu and input the characteristics
for the connected probe:
•
XTAL FREQUENCY
-
The frequency rating of the probe
•
EFF. DIAMETER
-
The effective diameter rating of the probe element
•
DELAY VELOCITY
-
User-determined delay-line velocity
Note:
These characteristics cannot be changed for any probes other than
PROBE #0
.