![GE Ultrasonic USM Go Operator'S Manual Download Page 127](http://html.mh-extra.com/html/ge/ultrasonic-usm-go/ultrasonic-usm-go_operators-manual_128490127.webp)
Chapter 3. Making Measurements
USM Go Operator’s Manual
109
3.16.2a Interpreting JISDAC Lines and Classes
The
JISDAC
display contains a recorded reference line and five fixed offset lines. Three of the lines are labeled as follows:
•
H Line
– Connects the recorded reference points
•
M Line
– Fixed offset positioned 6 dB below the H line
•
L Line
– Fixed offset positioned 12 dB below the H line
Any one of these three lines can be identified as the reference from which measurements are based (see the
BOLD LINE
function below). In
addition, the remaining three offset lines are drawn at 6, 12, and 18 dB above the H line.
The regions of the A-Scan display bordered by the H, M, and L lines are identified with JIS Class numbers. These identifying numbers can
also be displayed as results.
CLS
reports the region that contains the peak of the Gate A triggering echo.
Summary of Contents for Ultrasonic USM Go
Page 2: ......
Page 4: ... no content intended for this page proceed to next page ...
Page 18: ...Preface xvi USM Go Operator s Manual no content intended for this page proceed to next page ...
Page 162: ...144 USM Go Operator s Manual no content intended for this page proceed to next page ...
Page 178: ...Appendix C EN 12668 Specifications 160 USM Go Operator s Manual ...
Page 189: ......