![GE Ultrasonic USM Go Operator'S Manual Download Page 106](http://html.mh-extra.com/html/ge/ultrasonic-usm-go/ultrasonic-usm-go_operators-manual_128490106.webp)
Chapter 3. Making Measurements
88
USM Go Operator’s Manual
3.9.2
Working with TCG
In TCG mode the instrument uses the recorded reference points to calculate an amount of gain correction required to display each echo from
same-size reflectors at the same amplitude. The recorded reference point data is stored until replaced or edited. To use the stored reference
points and operate in
TCG
mode:
1.
Enter the
SETUP
menu, and then select the
TCG/DAC MODE
function.
2.
Activate the
TCG/DAC MODE
function, and select the
TCG
option.
3.
Select the
TCG DISPLAY
function, and turn this feature
ON
or
OFF
.
Note:
The TCG curve begins at the first reference point recorded. It then proceeds horizontally from the amplitude of this first reference
point to the depth (i.e. time position) of the last reference point recorded.
Note:
When rectification is set to
RF
, reference lines are not displayed.