Section 4 — Measurement Types and their Uses
69
Measuring Cross Channel Phase
The Cross Channel Phase measurement option has been enhanced. It’s
easier to change the speed of interest during the recording. This could
be the actual speed of the machine or the frequency of interest.
Additionally you can choose the required filter bandwidth you wish to
monitor, and the number of averages you wish to take. While recording,
a graph displays on your screen, showing your data. You can adjust the
position of the cursor on this graph to access more details about a peak
of interest, including phase and coherence data.
Setup
To select your preferred settings for a cross channel phase recording:
1.
Connect two sensors to the appropriate machine locations.
2.
Assign two sensors and enable both channels:
a.
Press
> Press
Sensor Setup
.
b.
Create or select your sensors as required.
c.
Press
to continue.
NOTE: The lowest channel will be used as the reference channel.
For example, if you have enabled CH2 and CH3, CH2 will be used
as the reference channel.
3.
Press
or
Measure.
4.
Use the navigation keys provided to locate and select
Cross
Channel Phase
> Press
.
5.
Press
Store Units
> Select ONE of these units according to
your requirements:
•
Acceleration
•
Velocity
•
Displacement
6.
Press
.
7.
Press
Speed/Freq of Interest
.
8.
Enter the required rpm value > Press
.
9.
Press
Filter Bandwidth
>
Select the required bandwidth.
Summary of Contents for SCOUT100-Ex
Page 18: ...Section 1 Introduction 7 Installation Control Drawing for Hazardous Locations ...
Page 20: ...Section 1 Introduction 9 Installation Control Drawing GE2214VBX for ATEX Hazardous Locations ...
Page 21: ...Section 1 Introduction 10 IECEx ATEX Declaration of Conformity ...
Page 40: ...Section 2 Instrument Fundamentals 29 Viewing hint labels Accessing more detailed help ...
Page 44: ...Section 2 Instrument Fundamentals 33 Using the advanced options ...
Page 54: ...Section 3 Sensors 43 Selecting the appropriate voltage for your tachometer and machine ...
Page 85: ...Section 4 Measurement Types and their Uses 74 Selecting the required parameter settings ...
Page 111: ...Section 4 Measurement Types and their Uses 100 Using the keypad to enter measurements ...
Page 116: ...Section 4 Measurement Types and their Uses 105 Example of tracked measurement ...
Page 120: ...Section 4 Measurement Types and their Uses 109 Saving multi channel measurements ...
Page 124: ...Section 5 Set Measurement Options 113 Spectral lines submenu ...
Page 164: ...Section 8 Send and Receive Information 153 Transferring recordings to Ascent ...
Page 194: ...Section 9 Working with Routes 183 Linear Speed ...
Page 241: ...Section 11 Balancing Rotors 230 Splitting a weight ...
Page 247: ...Section 12 Balancing Walkthrough 236 Taking an initial reading ...