Section 13 — General Setup Options
258
Overwrite Primary Database with Mirror Backup
Your instrument includes two operational databases:
•
A primary database used to store system information,
measurements, and setup data.
•
A secondary mirror backup database. The mirror database is an
exact duplicate of the current system database, and is
continually updated in real time.
If database corruption is detected (e.g. from a memory failure), your
instrument will automatically prompt you to overwrite the bad primary
database with the good mirror database when you switch it on.
To overwrite your current system database with the mirror backup
database:
1.
Press
Options
on the main menu.
2.
Press
Memory & System
to open the
Memory
screen.
3.
Press
Swap to Mirror DB
.
4.
Press
to overwrite the primary database with the mirror
backup, or
to cancel.
NOTE: Once the mirror database has been overwritten, you should
immediately import all data from your instrument to Ascent. You
should then format your memory card before taking any further
measurements.
Summary of Contents for SCOUT100-Ex
Page 18: ...Section 1 Introduction 7 Installation Control Drawing for Hazardous Locations ...
Page 20: ...Section 1 Introduction 9 Installation Control Drawing GE2214VBX for ATEX Hazardous Locations ...
Page 21: ...Section 1 Introduction 10 IECEx ATEX Declaration of Conformity ...
Page 40: ...Section 2 Instrument Fundamentals 29 Viewing hint labels Accessing more detailed help ...
Page 44: ...Section 2 Instrument Fundamentals 33 Using the advanced options ...
Page 54: ...Section 3 Sensors 43 Selecting the appropriate voltage for your tachometer and machine ...
Page 85: ...Section 4 Measurement Types and their Uses 74 Selecting the required parameter settings ...
Page 111: ...Section 4 Measurement Types and their Uses 100 Using the keypad to enter measurements ...
Page 116: ...Section 4 Measurement Types and their Uses 105 Example of tracked measurement ...
Page 120: ...Section 4 Measurement Types and their Uses 109 Saving multi channel measurements ...
Page 124: ...Section 5 Set Measurement Options 113 Spectral lines submenu ...
Page 164: ...Section 8 Send and Receive Information 153 Transferring recordings to Ascent ...
Page 194: ...Section 9 Working with Routes 183 Linear Speed ...
Page 241: ...Section 11 Balancing Rotors 230 Splitting a weight ...
Page 247: ...Section 12 Balancing Walkthrough 236 Taking an initial reading ...