Section 11 — Balancing Rotors
223
To specify your channel position:
1.
Press
Balance
.
2.
Press
Setup
.
3.
Press
# Sensors
.
4.
Select the number of sensors to use > Press
.
5.
Press
CH1 Position
> Enter the angle from the tachometer to
your sensor
> Press
.
6.
For dual plane balancing, repeat the above step if you want your
instrument to recommend trial weight placement angles for both
planes.
Using channel position
Averages
Averaging the out of balance values can increase the accuracy of your
measurements. This enables your instrument to make more precise
recommendations for the size and placement of balance weights. A
higher number of averages produces more accurate measurements, but
lengthens the recording time accordingly.
Summary of Contents for SCOUT100-Ex
Page 18: ...Section 1 Introduction 7 Installation Control Drawing for Hazardous Locations ...
Page 20: ...Section 1 Introduction 9 Installation Control Drawing GE2214VBX for ATEX Hazardous Locations ...
Page 21: ...Section 1 Introduction 10 IECEx ATEX Declaration of Conformity ...
Page 40: ...Section 2 Instrument Fundamentals 29 Viewing hint labels Accessing more detailed help ...
Page 44: ...Section 2 Instrument Fundamentals 33 Using the advanced options ...
Page 54: ...Section 3 Sensors 43 Selecting the appropriate voltage for your tachometer and machine ...
Page 85: ...Section 4 Measurement Types and their Uses 74 Selecting the required parameter settings ...
Page 111: ...Section 4 Measurement Types and their Uses 100 Using the keypad to enter measurements ...
Page 116: ...Section 4 Measurement Types and their Uses 105 Example of tracked measurement ...
Page 120: ...Section 4 Measurement Types and their Uses 109 Saving multi channel measurements ...
Page 124: ...Section 5 Set Measurement Options 113 Spectral lines submenu ...
Page 164: ...Section 8 Send and Receive Information 153 Transferring recordings to Ascent ...
Page 194: ...Section 9 Working with Routes 183 Linear Speed ...
Page 241: ...Section 11 Balancing Rotors 230 Splitting a weight ...
Page 247: ...Section 12 Balancing Walkthrough 236 Taking an initial reading ...