Section 4 — Measurement Types and their Uses
80
Long Time Waveforms
Long time waveforms enable you to take a continuous recording over a
long time period (several hours, if required). This is useful for capturing
transients. Once you have started measuring, recordings are taken
continuously until your specified time period has elapsed, or you stop the
measurement manually. The length of time you can record for depends
on the Equivalent Fmax. This determines the sample rate/lines of
resolution and the available memory remaining in the instrument.
Measure Long Time Waveforms
Duration is the total amount of time available for recording and is
displayed in seconds. You can record for the full length of time available
or set a shorter time period. Your chosen time period is shown on top,
while the available time is shown in brackets below this value.
The Equivalent Fmax (which determines the maximum frequency
displayed on your screen) will affect the available recording time. A high
Fmax reduces recording time, while a low Fmax increases the total
amount of time available for recording.
While your instrument is recording, you can display the signal on your
screen as a waveform. Alternatively, you can convert it to a spectral
display by pressing
Display Type & Resolution
and selecting your
preferred display type.
Summary of Contents for SCOUT100-Ex
Page 18: ...Section 1 Introduction 7 Installation Control Drawing for Hazardous Locations ...
Page 20: ...Section 1 Introduction 9 Installation Control Drawing GE2214VBX for ATEX Hazardous Locations ...
Page 21: ...Section 1 Introduction 10 IECEx ATEX Declaration of Conformity ...
Page 40: ...Section 2 Instrument Fundamentals 29 Viewing hint labels Accessing more detailed help ...
Page 44: ...Section 2 Instrument Fundamentals 33 Using the advanced options ...
Page 54: ...Section 3 Sensors 43 Selecting the appropriate voltage for your tachometer and machine ...
Page 85: ...Section 4 Measurement Types and their Uses 74 Selecting the required parameter settings ...
Page 111: ...Section 4 Measurement Types and their Uses 100 Using the keypad to enter measurements ...
Page 116: ...Section 4 Measurement Types and their Uses 105 Example of tracked measurement ...
Page 120: ...Section 4 Measurement Types and their Uses 109 Saving multi channel measurements ...
Page 124: ...Section 5 Set Measurement Options 113 Spectral lines submenu ...
Page 164: ...Section 8 Send and Receive Information 153 Transferring recordings to Ascent ...
Page 194: ...Section 9 Working with Routes 183 Linear Speed ...
Page 241: ...Section 11 Balancing Rotors 230 Splitting a weight ...
Page 247: ...Section 12 Balancing Walkthrough 236 Taking an initial reading ...