Chapter 3. Using the Settings Menu
46
moisture.IQ User’s Manual
3.8 Fault Alarm Setup
The
Fault Alarm Setup
menu (see
below) allows you to specify how the moisture.IQ dedicated fault
alarm relay responds to a fault condition and it lists the source of any current fault conditions.
Figure 40: The Fault Alarm Setup Screen
•
Tap the
In Case of Fault
button to toggle between the
Energized
and
DeEnergized
options. This selection
determines whether the relay is
Energized
or
DeEnergized
(fail-safe operation) when a fault condition is
detected. The source of a fault is displayed in the fault list in
above.
Note:
The fault alarm relay is tripped whenever any sensor is open, shorted or disconnected. It will also trip
when an overheating condition is detected.
•
After completing your selection, either tap the
Enter
button to save your changes, or tap the
Cancel
button to discard the changes and keep the original settings. In either case, you will be returned to the
Settings Menu
.
Summary of Contents for moisture.IQ
Page 1: ...Moisture and Oxygen 910 298 Rev E September 2017 bhge com moisture IQ User s Manual ...
Page 2: ......
Page 4: ... no content intended for this page ...
Page 12: ...Preface x moisture IQ User s Manual no content intended for this page ...
Page 45: ...moisture IQ User s Manual 33 Chapter 2 Operation ...
Page 49: ...moisture IQ User s Manual 37 Chapter 2 Operation ...
Page 50: ...Chapter 2 Operation 38 moisture IQ User s Manual ...
Page 110: ...Chapter 8 Troubleshooting 98 moisture IQ User s Manual no content intended for this page ...
Page 180: ... no content intended for this page ...
Page 181: ......