Certification & Safety Statements for the moisture.IQ Moisture and Oxygen Analyzer
Special Conditions for Safe Use
XP conditions of use for ATEX/IECEx and Zones:
•
The flameproof joints of the equipment are not intended to be repaired. Consult the manufacturer if repair of the flameproof
joints is necessary.
•
Consult the manufacturer for genuine replacement enclosure cover fasteners.
Installation
When installing this apparatus, the following requirements must be met:
•
The temperature rating of the field wiring must be 70°C or greater.
•
XP version unused entries should be closed with suitably certified blanking elements.
•
The installation should comply with controlled drawing 752-364 (see
) and drawing
752-513 Sheets 1-7
(see
through
•
The installation should comply with configuration drawing 712-1889 Sheets 1-3 (see
through
•
The installation should comply with installation drawings 712-2126 stainless steel weather-proof (see
), 712-2127 explosion-proof (see
), and 712-2130 fiberglass weather-proof (see
•
The product cannot be repaired by the user. It must be replaced by an equivalent certified product. Repairs should only be
carried out by the manufacturer or by an approved repairer.
•
Only trained, competent personnel may install, operate and maintain the equipment.
•
If you have questions about whether the apparatus can be used safely in the intended area under the expected operating
conditions, please call a BHGE applications engineer.
•
See the back cover of the
User’s Manual
for contact information.
Summary of Contents for moisture.IQ
Page 1: ...Moisture and Oxygen 910 298 Rev E September 2017 bhge com moisture IQ User s Manual ...
Page 2: ......
Page 4: ... no content intended for this page ...
Page 12: ...Preface x moisture IQ User s Manual no content intended for this page ...
Page 45: ...moisture IQ User s Manual 33 Chapter 2 Operation ...
Page 49: ...moisture IQ User s Manual 37 Chapter 2 Operation ...
Page 50: ...Chapter 2 Operation 38 moisture IQ User s Manual ...
Page 110: ...Chapter 8 Troubleshooting 98 moisture IQ User s Manual no content intended for this page ...
Page 180: ... no content intended for this page ...
Page 181: ......