moisture.IQ User’s Manual
9
Chapter 1. Installation
1.8 Installing the Probes
After you mount the sample system, you must insert the moisture probes into the sample cells. In addition, you
must check, prepare, and connect any installed oxygen cells to the gas line.
1.8.1 Moisture Probes
BHGE
Moisture Image Series
(MIS) and
M-Series
moisture probes have 3/4 inch-16 straight threads, with an
O-ring seal, to the secure probes into the sample system or directly into the process line. Other fittings are
available for special applications.
Moisture probes are usually installed into a sample system. The sample system protects the probes from
coming into contact with damaging elements in the process. Moisture probes are installed into a cylindrical
shaped container, called the
sample cell
, which is included as part of your sample system. The sample cell is
identified by a label on the sample system plate.
To install a moisture probe that has a 3/4 inch-16 straight thread, insert it into the sample cell so it is
perpendicular to the sample inlet. Screw the probe into cell, making sure not to cross thread it, and tighten it
securely.
below shows a typical moisture probe installation, with the probe mounted in a BHGE
sample cell. Install any moisture probes with different fittings in the appropriate manner.
Figure 4: Typical Moisture Probe in Sample Cell
Note:
Standard moisture probes have a sintered stainless steel shield that protects the aluminum oxide sensor.
Be sure to leave the shield in place for maximum protection.
It is important to eliminate all gas or liquid leaks for safety reasons and to ensure that measurements are not
affected by ambient contamination. For gas applications, you should check for leaks using a soap bubble
solution.
IMPORTANT:
Refer to the
Calibration Data Sheets
to connect the probes to their corresponding channels. If a
probe is connected to the wrong channel, the meter will display incorrect data.
CAUTION!
If mounting the moisture probes directly into the process line, you must consult BHGE for
proper installation instructions and precautions.
Probe
Sample Cell
Inlet
Outlet
Summary of Contents for moisture.IQ
Page 1: ...Moisture and Oxygen 910 298 Rev E September 2017 bhge com moisture IQ User s Manual ...
Page 2: ......
Page 4: ... no content intended for this page ...
Page 12: ...Preface x moisture IQ User s Manual no content intended for this page ...
Page 45: ...moisture IQ User s Manual 33 Chapter 2 Operation ...
Page 49: ...moisture IQ User s Manual 37 Chapter 2 Operation ...
Page 50: ...Chapter 2 Operation 38 moisture IQ User s Manual ...
Page 110: ...Chapter 8 Troubleshooting 98 moisture IQ User s Manual no content intended for this page ...
Page 180: ... no content intended for this page ...
Page 181: ......