moisture.IQ User’s Manual
11
Chapter 1. Installation
1.8.3 Adapting an Existing M-Series Probe Cable
If an existing moisture measurement system uses a
M-Series
moisture probe, the cable for that probe may be
adapted for use with an
MISP2
probe. To retrofit the existing system, refer to
below and complete
the following steps:
Figure 6: M-MISP2 Converter Cable
1.
Locate the existing
M-Series
probe and disconnect the cable from the probe.
2.
Replace the
M-Series
probe with an
MISP2
probe.
3.
Using BHGE
adapter cable 704-1649-00
, connect one end of the cable to the
MISP2
probe and connect
the other end to the existing
M-Series
cable connector.
4.
At the moisture analyzer end of the
M-Series
cable, locate the same
M-Series
cable and disconnect all
flying leads from the pluggable connector.
5.
Locate the
GREEN
and
RED
wires of the
M-Series
cable and connect them to the
BLACK
and
WHITE
wires
at the
MISP2
probe gray pluggable connector. (The
GREEN
wire is connected to the
MISP
connector
"BLK"
label and
RED
wire is connected to the
MISP
connector "
WHT
" label.) See
Probes (MISP and MISP2)” on page 19
for help.
6.
Trim all other flying leads from the
M-Series
cable flush with the jacket.
7.
Plug the newly-wired
MISP2
probe pluggable connector into the moisture.IQ.
MISP2 Connector
[Total Cable Length = 6.50”]
M Series Connector
Summary of Contents for moisture.IQ
Page 1: ...Moisture and Oxygen 910 298 Rev E September 2017 bhge com moisture IQ User s Manual ...
Page 2: ......
Page 4: ... no content intended for this page ...
Page 12: ...Preface x moisture IQ User s Manual no content intended for this page ...
Page 45: ...moisture IQ User s Manual 33 Chapter 2 Operation ...
Page 49: ...moisture IQ User s Manual 37 Chapter 2 Operation ...
Page 50: ...Chapter 2 Operation 38 moisture IQ User s Manual ...
Page 110: ...Chapter 8 Troubleshooting 98 moisture IQ User s Manual no content intended for this page ...
Page 180: ... no content intended for this page ...
Page 181: ......