Chapter 9. Specifications
9.7 Output Specifications
Analog
Six per 3-CH module, two per 1-CH module
Internally isolated
12 bit (0.025% Resolution)
Standard Switch-Selectable Outputs
•
0 to 2 V, 10K ohm minimum load resistance
•
0 to 20 mA, 400 ohm maximum series resistance
•
4 to 20 mA, 400 ohm maximum series resistance
Each output can correspond to any one parameter on that channel. Zero and Span are user-programmable
within the range of the instrument and the corresponding sensor.
Digital Outputs
•
Choice of
RS232
or
RS485
serial communications port. Available
baud rates
include:
9600, 19200, 38400, 57600, 115200
•
USB A:
Compliant to
USB 1.1
, master, host for USB storage
•
USB B:
Compliant to
USB 1.1
, slave (factory use only)
•
Ethernet:
RJ-45, supports
10BASE-T
and
100BASE-TX
Alarm Relays
Six per 3-CH module, two per 1-CH module. 1 Form C relay SPDT, rated for 2 amps at 28 VDC/28 VAC. Available
for high and low limits. The relay contacts can be set to trip at any numerical level within the range of the
instrument.
Output Updating
The microprocessor samples, processes data, and calculates values for each channel sequentially. The
minimum update time is 1 second depending on configuration and mode. Channels are updated sequentially.
Summary of Contents for moisture.IQ
Page 1: ...Moisture and Oxygen 910 298 Rev E September 2017 bhge com moisture IQ User s Manual ...
Page 2: ......
Page 4: ... no content intended for this page ...
Page 12: ...Preface x moisture IQ User s Manual no content intended for this page ...
Page 45: ...moisture IQ User s Manual 33 Chapter 2 Operation ...
Page 49: ...moisture IQ User s Manual 37 Chapter 2 Operation ...
Page 50: ...Chapter 2 Operation 38 moisture IQ User s Manual ...
Page 110: ...Chapter 8 Troubleshooting 98 moisture IQ User s Manual no content intended for this page ...
Page 180: ... no content intended for this page ...
Page 181: ......