moisture.IQ User’s Manual
73
Chapter 5. Using the Configuration Menu
5.4.4 Entering Saturation Constants
Henry's Law
is applicable to
ppmw
moisture measurements in all organic liquids. It states that, “
At a constant
temperature, the amount of a given gas that dissolves in a given type and volume of liquid is directly proportional
to the partial pressure of that gas in equilibrium with that liquid.”
In other words,
PPMw = (Pw/Ps) x Cs
. To
calculate ppmw moisture values for an organic liquid with the moisture.IQ,
saturation values
(
Cs
), as a function
of temperature, must be entered in the
Cs Table
shown in
Figure 64: Cs Table (Henry’s Law) Screen
To enter the Cs values for your application, complete the following steps:
1.
Tap the
Channel
button and specify the channel for which the Cs curve should be applied.
2.
Use the
Insert Row
and
Remove Row
buttons to configure the table with up to 10 rows.
3.
To enter or edit the data in the table, tap the desired table cell to highlight it in blue. Then, tap the
Edit
Value
button to enter the data for that cell.
Figure 65: Selecting the Chemical
Summary of Contents for moisture.IQ
Page 1: ...Moisture and Oxygen 910 298 Rev E September 2017 bhge com moisture IQ User s Manual ...
Page 2: ......
Page 4: ... no content intended for this page ...
Page 12: ...Preface x moisture IQ User s Manual no content intended for this page ...
Page 45: ...moisture IQ User s Manual 33 Chapter 2 Operation ...
Page 49: ...moisture IQ User s Manual 37 Chapter 2 Operation ...
Page 50: ...Chapter 2 Operation 38 moisture IQ User s Manual ...
Page 110: ...Chapter 8 Troubleshooting 98 moisture IQ User s Manual no content intended for this page ...
Page 180: ... no content intended for this page ...
Page 181: ......