moisture.IQ User’s Manual
17
Chapter 1. Installation
1.9.3.1 M-Series Probes
M-Series probes are primarily used for moisture measurement, but they can be used to measure temperature
as well. If ordered, an optional temperature thermistor is included in the moisture probe and requires an
additional connection.
If the M-Series probe does not have a thermistor, then you may use a two-wire shielded cable with a
bayonet-type connector to connect the probe to the electronics unit. If an optional thermistor is installed, then
use a four-wire shielded cable with a bayonet-type connector to connect the M-Series Probe to the electronics
unit. The M-Series Probe may be located up to 600 m (2000 ft) from the moisture.IQ.
Before making the electrical connections, connect the cable to the probe by inserting the bayonet-type
connector onto the probe and twisting the shell clockwise until it snaps into a locked position (approximately
1/8 of a turn). Wire the probe cable per the wiring diagram in
below. The M-Series terminal block
on the back panel of the moisture.IQ (see
) also notes the corresponding color from the
probe cable.
Figure 12: M-Series Probe Cable Wiring Diagram
Note:
The label
SHD
above corresponds to the
SH
label on the connector, and these terminals are
used to terminate the cable shield.
5
4
3
2
1
M-SERIES
BLACK
WHITE
SHD
GREEN
RED
M-SERIES
Temp
Moisture
.
Summary of Contents for moisture.IQ
Page 1: ...Moisture and Oxygen 910 298 Rev E September 2017 bhge com moisture IQ User s Manual ...
Page 2: ......
Page 4: ... no content intended for this page ...
Page 12: ...Preface x moisture IQ User s Manual no content intended for this page ...
Page 45: ...moisture IQ User s Manual 33 Chapter 2 Operation ...
Page 49: ...moisture IQ User s Manual 37 Chapter 2 Operation ...
Page 50: ...Chapter 2 Operation 38 moisture IQ User s Manual ...
Page 110: ...Chapter 8 Troubleshooting 98 moisture IQ User s Manual no content intended for this page ...
Page 180: ... no content intended for this page ...
Page 181: ......