Chapter 1. Installation
20
moisture.IQ User’s Manual
1.9.3.2 Moisture Image Series Probes (cont.)
If the
Moisture Image Series
probe requires assembly, complete that task before proceeding. Then, connect the
probe cable to the terminal block labeled
MIS
on the back panel of the electronics unit (see
You can connect the
Moisture Image Series Probe
to any channel. However, if you are also using other sensors
such as M-Series probes, be sure to connect the
Moisture Image Series
probe to an unassigned channel.
IMPORTANT:
Check the
Calibration Data Sheets
of all installed sensors to determine which channels already
have probes assigned to them.
Figure 16: MIS Probe Connections
Note:
Additional MIS probes may be connected to
CH4
,
CH5
and
CH6
in the
Module B
connector group on the
right side of the back panel.
After you make the back panel connections, connect the other end of the probe cable to the
Moisture Image
Series
probe per the wiring diagram in
Trim all other flying leads from the cable flush with the
jacket.
After you complete the
Moisture Image Series
probe connections, you must activate the probe on the installed
channel as described in
“The Probe Configuration Screen” on page 60
Figure 17: MIS Probe Cable Wiring Diagram
Module A - CH2
MIS Probe Connections
Module A - CH3
Module A - CH1
MISP 1
or MISP 2
WHITE
BLACK
2
1
MIS
Summary of Contents for moisture.IQ
Page 1: ...Moisture and Oxygen 910 298 Rev E September 2017 bhge com moisture IQ User s Manual ...
Page 2: ......
Page 4: ... no content intended for this page ...
Page 12: ...Preface x moisture IQ User s Manual no content intended for this page ...
Page 45: ...moisture IQ User s Manual 33 Chapter 2 Operation ...
Page 49: ...moisture IQ User s Manual 37 Chapter 2 Operation ...
Page 50: ...Chapter 2 Operation 38 moisture IQ User s Manual ...
Page 110: ...Chapter 8 Troubleshooting 98 moisture IQ User s Manual no content intended for this page ...
Page 180: ... no content intended for this page ...
Page 181: ......