GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
FUNCTIONAL CHECKS
4–30
REV 5
4–3–3 Utility Menu (Continued)
(I) MEDIA INITIALIZE
This initialize the floppy disk.
Type 9 and then press Return key when the Utility Menu is displayed. The Media Initialize screen is displayed as
shown in ILLUSTRATION 4–23.
MEDIA INITIALIZATION
ILLUSTRATION 4–23
Procedure for Changing Each Settings
Insert an 128MB MO (Magneto Optical) disk (or 230MB disk for the conosle with software version 3.10 or later)
and press the Return key. The system automatically starts initializing the MO disk. It takes approximately 15
minutes to complete initializing a disk.
Note
The 230MB MO disk is available on the console with software version 3.10 or later.
During the system is initializing MO disk, “Initializing” is displayed on the CRT monitor. After initializing, “Suc–
ceeded” or “Failed” is displayed on the CRT monitor. Then, the message “Press Y to continue, C R to
quit” is displayed. Type Y and press Return to continue to use this function. Press and hold Control, and press R
to quit this function.