GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
TM
8–21
OPTIONS
REV 5
8–2–10 Operational Check-out (Continued)
3.
Power the Video Cassette Recorder ON.
4.
Press New Patient key when the PATIENT ENTRY MENU is displayed on the CRT monitor screen of
LOGIQ
400.
5.
Set the system so that the Video Cassette Recorder can work by the key operation on the console.
The following steps (step a. to k.) indicates one example when you set the VCR control with the following condi-
tions:
I. Use S-Video cable for connecting between LOGIQ
400 and VCR.
II. Connect RS232C cable with “port A” located at rear panel of LOGIQ
400.
a. Press the Top Menu Select key located on the keyboard. The sub–menu selected previously is displayed on
the CRT monitor screen.
SOFTWARE MENU CONTROL KEYS
ILLUSTRATION 8–15
b. Press the Top Menu Select key again. The top menu is displayed on CRT monitor screen as shown in
ILLUSTRATION 8–16.
TOP MENU
ILLUSTRATION 8–16
c. Select the Set Up Menu using the
or
key of Sub Menu Select keys. The Set Up selection is dis-
played in reverse video as shown in ILLUSTRATION 8–16.