GE MEDICAL SYSTEMS
LOGIQ 400 SERVICE MANUAL
2127661
RENEWAL PARTS
6–78
REV 9
6–2–5 CRT Cap Set (FRU No. 108)
Time Required
1 Minute
Tool Required
Not necessary
Procedure
Refer to ILLUSTRATION 6–21.
1.
Turn OFF the system.
2.
Remove the Escutcheon Front Door (FRU 107). Refer to 6–2–4 on page 6–77.
3.
Pull out two CRT Caps from the front side of the Monitor Assy.
CRT CAP SET REMOVAL
ILLUSTRATION 6–21
Note
The CRT Cap Set (FRU108) includes the key top of lamp switch.