![GE LOGIQ 400 Service Manual Download Page 108](http://html.mh-extra.com/html/ge/logiq-400/logiq-400_service-manual_128623108.webp)
GE MEDICAL SYSTEMS
2127661
LOGIQ 400 SERVICE MANUAL
FUNCTIONAL CHECKS
4–44
REV 6
4–4–2 Power Supply Adjustment Procedure (continued)
TABLE 4–5
POWER SUPPLY 3 MEASUREMENT LOCATION
POWER SUPPLY
MEASURE AT
RETURN AT
+5V for Digital
+5VD Terminal on Power Supply
See ILLUSTRATION 4–33
GND Terminal on Power Supply
See ILLUSTRATION 4–33
+5V for Analog
+5VA Terminal on Power Supply
See ILLUSTRATION 4–33
GND Terminal on Power Supply
See ILLUSTRATION 4–33
+12V
+12V Terminal on Power Supply
See ILLUSTRATION 4–33
GND Terminal on Power Supply
See ILLUSTRATION 4–33
–5V
–5V Terminal on Power Supply
See ILLUSTRATION 4–33
GND Terminal on Power Supply
See ILLUSTRATION 4–33
+15V
+15V Terminal on Power Supply
See ILLUSTRATION 4–33
GND Terminal on Power Supply
See ILLUSTRATION 4–33
TABLE 4–6
POWER SUPPLY 3 MEASUREMENT TOLERANCES
POWER SUPPLY
MIN.
MAX.
+5V for Digital
+5.25V
+5.35V
+5V for Analog
+5.35V
+5.45V
+12V
+12.10V
+12.20V
–5V
–5.10V
–5.20V
+15V
+15.40V
+15.50V